OLR Series 1000 OnLine Rheometer
The OnLine Rheometer (OLR) is designed to continuously measure, plot, and report the rheological properties of drilling, completion, and fracturing fluids in the pipe. The OLR uses a squeezeflow technique; it measures storage and loss moduli by imposing a small cyclic deformation on a liquid sample at a variety of frequencies, from 1-100 Hz. Quality control parameters and user interface are based on ?* (complex viscosity [Pa s]), as measured by the OLR. The response of the liquid is measured and displayed in terms of G' (storage modulus [Pa]), G'' (loss modulus [Pa]), d (phase angle [Degrees]), and ?*. Features Fast and Reliable Real Time Results Low Maintenance Out of Specification Diagnosis Plug-and-Play Installation Precise and Robust Sensor System Process Control Process Monitoring Quality Control State of the Art Technology Time and Cost Savings Specifications Frequency Range: 1 - 100 Hz Viscosity Measurement Range (geometry dependent): 0.25 - 200,000 Pa s Force Range: 0 - 44.48 N Operating Range: Temperature: 14° - 230°F (-10° - 110°C) Pressure: 0.7 - 145 psi (4.8 - 1,000 kPa) Temperature Measurement: RTD PT100 Ohm, Class A Size: 11.6" × 18.1" × 23.8" (30 × 46 × 61 cm) Weight: 145 lb (65 kg) Requirements Electrical: 115 - 230 VAC, 50/60 Hz Power Input Maximum: 10 W Fuses: 5A Suitable for 1" to 8" pipe (DN25 - DN200) Software (SOLR) Features Set-up of process monitoring and quality control parameters Monitoring and control of process either through SOLR or through the factory PLC Out-of-specification product diagnosis